Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×

C
510(k) Premarket Notification Analysis of FDA Recall Data

William H. Maisel, MD, MPH

EXECUTIVE SUMMARY

The 510(k) process requires a device manufacturer to notify the Food and Drug Administration (FDA) before it intends to market a device and to establish that the device is “substantially equivalent” to a legally marketed “predicate” device that does not require premarket approval (PMA). A recall is an action taken to address a problem with a medical device that violates FDA law. Recalls occur when a medical device is defective and/or when it could be a risk to health.

Analysis of FDA’s 510(k) Database (1996–2009) and Recall Database (2003–2009) revealed the following:

  1. 48,402 510(k)s were cleared by FDA between January 1, 1996, and December 31, 2009, and were available for analysis.

  2. From 2003 to 2009, 3,132 unique 510(k)s were subject to recall. Among 510(k)s affected by recall, 73.9% were recalled a single time and 26.1% were recalled more than once, including nearly 2% that were recalled more than five times.

  3. Among 510(k)s cleared in 2003–2009, 98.4% remained free of recall 1 year following the decision. Longer-term follow-up shows that 92.6% and 91.5% of 510(k)s remain free of recall 5 and 6 years, respectively, following regulatory clearance.

  4. The annual 510(k) recall rate is highest in the first 3 years following clearance (1.6–1.9%/year). Lower recall rates are observed in years 5 and 6 post clearance (0.9–1.1%/year).

  5. More than half the 510(k) recalls are due to manufacturing process

Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×

errors (28.8%) or device design issues (28.4%). Materials and component issues (16.3%) and change control processes (11.9%) account for the majority of the remaining 510(k) recalls.

  1. Compared to 510(k)s unaffected by recall, recalled 510(k)s are more likely to have been reviewed by a third party or submitted as a Special application (rather than Traditional or Abbreviated). Recalls are also more likely to affect 510(k)s involving life sustaining devices and Class III devices.

OVERVIEW OF 510(k) PREMARKET NOTIFICATION PROGRAM AND FDA RECALLS

The 510(k) process requires a device manufacturer to notify FDA before it intends to market a device and to establish that the device is “substantially equivalent” to a legally marketed “predicate” device that does not require a PMA.1

One measure of the success of the 510(k) premarket notification process would be to assess the performance and reliability of the thousands of individual devices that have been cleared via this pathway. This is both impractical and impossible. Recalls and Medical Device Reports represent surrogate markers of device reliability.

A recall is an action taken to address a problem with a medical device that violates FDA law.2 Recalls occur when a medical device is defective and/or when it could be a risk to health. Recalls may be conducted on a firm’s own initiative, by FDA request, or by FDA order under statutory authority. Importantly, recalls may be issued when there is only a small chance (sometimes <1%) of device malfunction or patient injury. Recalls are classified by FDA as follows:

  • Class I recall: a situation in which there is a reasonable probability that the use of or exposure to a violative product will cause serious adverse health consequences or death.

  • Class II recall: a situation in which use of or exposure to a violative product may cause temporary or medically reversible adverse health consequences or where the probability of serious adverse health consequences is remote.

  • Class III recall: a situation in which use of or exposure to a violative product is not likely to cause adverse health consequences.

1

Government Accountability Office. Medical Devices: FDA Should Take Steps to Ensure That High-Risk Device Types Are Approved Through the Most Stringent Premarket Review Process. January 2009. Accessed June 16, 2009 at http://www.gao.gov/new.items/d09190.pdf.

2

U.S. Food and Drug Administration. Recalls, Market Withdrawals, and Safety Alerts. Accessed at http://www.fda.gov/Safety/Recalls/ucm165546.htm.

Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×

Medical Device Reporting (MDR) is a mechanism by which FDA receives information on significant medical device adverse events from manufacturers, importers, and user facilities (hospitals, nursing homes, etc.). Because of incomplete adverse event descriptions, significant underreporting of device-related adverse events, and the absence of denominator data, these data have significant limitations. Nevertheless, MDR analysis can provide some insights into the performance of the 510(k) program.

METHODOLOGY

Methods 2.1

Two primary databases—510(k) database and recall database—were used to conduct the analysis presented in this report. A third database, containing MDR data, was also analyzed. Each database was provided by FDA to the Institute of Medicine. Data analysis was conducted independently of FDA.

Methods 2.1.1 510(k) Database for Years 1996–2009

Consisted of all 510(k) applications that were submitted between January 1, 1996, and December 31, 2009, and were found to be substantially equivalent (SE).

Methods 2.1.2 FDA Recall Database for Years 2003–2009

Consisted of all medical device recalls from January 1, 2003, to December 31, 2009.

Methods 2.1.3 Medical Device Reporting (MDR) Data

Consisted of a subset of all submitted MDRs. A subset of the MDRs occurring between January 1, 2003, and December 31, 2009, were analyzed.

Methods 2.2

All statistical calculations were performed using SAS (Version 9.1, Cary, NC). A two-sided P value of <0.05 was considered to be statistically significant. Chi-square tests, Mantel–Haenszel chi-square tests, and log-rank tests were used where appropriate.

Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×

DATA ANALYSIS

FIGURE C-1 Annual number of 510(k) applications.

FIGURE C-1 Annual number of 510(k) applications.

  • From 1996 to 2009, 48,402 510(k)s were available for analysis.

  • The annual number of applications ranged from a high of 4,286 in 1996 to a low of 2,890 in 2009.

Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
FIGURE C-2 Advisory committee assignments for submitted 510(k)s.

FIGURE C-2 Advisory committee assignments for submitted 510(k)s.

  • All devices submitted for clearance under the 510(k) program are given an Advisory Committee assignment.

  • Devices cleared via the 510(k) program are most often assigned to the General & Plastic Surgery (11.8%), Orthopedic (11.4%), General Hospital (10.2%), and Cardiovascular (9.8%) Advisory Committees.

  • 510(k) devices are least often assigned to the Ear, Nose, & Throat (1.5%), Hematology (1.8%), and Clinical Toxicology (1.9%) Advisory Committees.

  • While Advisory Committee assignments inform about the type of devices cleared, the committees are rarely involved in the premarket 510(k) process.

Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
FIGURE C-3 510(k) device classification.

FIGURE C-3 510(k) device classification.

  • More than 80% of 510(k) devices are classified as Class II, ~10% as Class I, and <2% as Class III.

  • Minor changes in the distribution of device classification have occurred from 1996 to 2002 compared to 2003–2009. Specifically, there has been a small reduction in the number of 510(k) devices classified as Class I, and a small increase in the number of Class II and Class III designations.

Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
FIGURE C-4 510(k) type, 1996–2009.

FIGURE C-4 510(k) type, 1996–2009.

  • The majority of 510(k)s are Traditional (80.9%). Fewer are Special (16.0%) or Abbreviated (3.0%).

  • Special 510(k)s are submitted when a manufacturer makes modifications to its own device, design control processes are appropriate, and design validation is performed. Abbreviated 510(k)s are submitted when FDA guidance, a special control, or recognized standard exists and the manufacturer intends to use it. All other 510(k) devices utilize the traditional pathway.

Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
FIGURE C-5 510(k) implantable and life-sustaining features, 2003–2009.

FIGURE C-5 510(k) implantable and life-sustaining features, 2003–2009.

  • Nearly one-quarter (23.4%) of 510(k)s are implantable devices.

  • The vast majority of 510(k) devices are NOT considered life-sustaining. However, a small percentage (3.5%) are life-sustaining devices.

Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
FIGURE C-6 Annual number of unique 510(k)s affected by recall.

FIGURE C-6 Annual number of unique 510(k)s affected by recall.

  • From 2003 to 2009, 3,132 unique 510(k)s were subject to FDA recall.

  • The annual number of recalled 510(k)s ranged from a low of 377 in 2004 to a high of 508 in 2008.

Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×

TABLE C-1 Number of Recalls per 510(k)

Recalls per 510(k)

# Occurrences

Percentage of 510(k) Recalls

1

2,298

73.9

2

492

15.8

3

150

4.8

4

74

2.4

5

37

1.2

6

16

0.5

7

15

0.5

8

10

0.3

9

8

0.3

10

2

0.1

>10

7

0.2

  • Among 510(k)s affected by recall, 73.9% were recalled a single time.

  • 26.1% of recalled 510(k)s were recalled more than once, including nearly 2% that were recalled more than five times.

  • Importantly, multiple recalls may be due to expansion of an initial recall to additional products with the same potential defect and do not necessarily represent multiple modes of product failure. Additionally, because some 510(k)s contain more than one device, multiple recalls do not necessarily represent product defects repeatedly affecting the same device.

Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
FIGURE C-7 Year of 510(k) decision for recalls occurring in 2003–2009.

FIGURE C-7 Year of 510(k) decision for recalls occurring in 2003–2009.

  • The original decision year for 510(k)s affected by recall in 2003–2009 is displayed.

  • Nearly half (49.9%) of the 510(k) recalls occurring in 2003–2009 were for products cleared in 1996–2002.

  • 40.6% of recalls were for products that were cleared in 2003–2009.

Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
FIGURE C-8 Recall-free 510(k) “survival,” 2003–2009.

FIGURE C-8 Recall-free 510(k) “survival,” 2003–2009.

  • Only 510(k)s cleared from January 1, 2003, to December 31, 2009, are included in this graph. Kaplan–Meier “survival” estimates were calculated using standard methodology.

  • Recall-free survival estimates demonstrate that 98.4% of 510(k)s cleared in 2003–2009 remained free of recall 1 year following the decision.

  • Longer term follow-up shows that 92.6% and 91.5% of 510(k)s remain free of recall 5 and 6 years, respectively, following regulatory clearance.

Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
FIGURE C-9 Annual 510(k) recall rate based on years since decision.

FIGURE C-9 Annual 510(k) recall rate based on years since decision.

  • The annual rate of recall for 510(k)s cleared January 1, 2003, to December 31, 2009, is displayed.

  • The annual recall rate is highest in the first 3 years following clearance (1.6–1.9%).

  • The lowest recall rates are observed in years 5 and 6 post clearance (0.9–1.1%).

Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×

TABLE C-2 Causes of 510(k) Recalls

Cause

Description

Recalls (%)

Manufacturing Process

Manufacturing process inadequately controlled, inadequate environmental controls, storage, packaging, labeling, equipment maintenance, material removal, etc.

28.8

Device Design

Failure of device to perform as intended despite meeting design specifications.

28.4

Materials/Components

Materials/components that are non-conforming, contaminated, degraded, counterfeit, or inadequately tested

16.3

Change Control

A change made to a specification, program, procedure, vendor, etc. that adversely affects a component, finished device, packaging, leveling, etc.

11.9

Employee Error

Employee error (not a systematic problem). Usually corrected by retraining.

7.1

Miscellaneous

 

7.5

TOTAL

 

100

  • FDA classifies the cause of each recall based on the available information.

  • More than half the 510(k) recalls are due to manufacturing process errors (28.8%) or device design issues (28.4%).

  • Materials and component issues (16.3%) and change control processes (11.9%) account for the majority of the remaining recalls.

  • Employee error is a less common cause of 510(k) recalls (7.1%).

Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
FIGURE C-10 Impact of predicate number on 510(k) recalls, 2004–2009.

FIGURE C-10 Impact of predicate number on 510(k) recalls, 2004–2009.

  • The number of predicates cited by each individual 510(k) is displayed.

  • Most 510(k)s utilize 1-5 predicates (>80%). Fewer than 5% utilize more than 10 predicates.

  • Fewer predicates (1-5) are associated with a lower rate of recall while a higher number of predicates (6-10, for example) are associated with an increased 510(k) recall rate.

  • Many of the submissions that cite multiple predicates contain multiple products (for example, submissions to the Office of In Vitro Diagnostic Device Evaluation and Safety [OIVD] for diagnostic tests). Therefore, one cannot conclude from these data that multiple predicates are unsafe without indexing the recall rate to the number of unique products at risk. This latter analysis could not be performed based on the available data.

Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
FIGURE C-11 Impact of predicate age on 510(k) recalls, 2004–2009.

FIGURE C-11 Impact of predicate age on 510(k) recalls, 2004–2009.

  • The age of the newest predicate cited for 510(k) clearance was <5 years in most (>75%) cases and <10 years in >90%.

  • The age of the oldest predicate was <5 years in ~ 50%, 6-10 years in ~ 25%, and >10 years ~25% of the time.

  • The newest predicate and the oldest predicate tended to be younger among recalled 510(k)s than among those 510(k)s that had been unaffected by recall.

Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
FIGURE C-12 Impact of 510(k) type on recall rate, 2003–2009.

FIGURE C-12 Impact of 510(k) type on recall rate, 2003–2009.

  • Traditional 510(k) were most common for both 510(k)s affected and unaffected by recalls.

  • Special 510(k)s represented a higher percentage of 510(k) type among recalled 510(k)s than among 510(k)s unaffected by recall (34.2% vs 22.3%).

  • Abbreviated 510(k)s were infrequent for both 510(k)s affected and unaffected by recall.

Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
FIGURE C-13 Association of 510(k) features with recall rate, 2003–2009.

FIGURE C-13 Association of 510(k) features with recall rate, 2003–2009.

  • 510(k)s affected by recall were less often cleared by the Office of Device Evaluation (67.9% vs 77.0%) and more often cleared by the OIVD (32.1% vs 23.0%) than devices unaffected by recall.

  • 510(k)s were more often cleared via third party review among recalled 510(k)s than among those unaffected by recall (9.9% vs 7.3%).

  • 510(k)s affected by recall were more often life-sustaining devices than 510(k)s unaffected by recall (8.5% vs 3.2%).

Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
FIGURE C-14 Device classification and 510(k) recall rate, 2003–2009.

FIGURE C-14 Device classification and 510(k) recall rate, 2003–2009.

  • More than 85% of 510(k)s were Class II devices for devices both affected and unaffected by recall.

  • There was a more than 3-fold increase in the percentage of Class III devices among 510(k)s affected by recall compared to those unaffected (5.6% vs 1.7%). In contrast, there was a reduction in the percentage of Class I devices among recalled 510(k)s compared to 510(k)s unaffected by recall (5.6% vs 1.7%).

Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
FIGURE C-15 Association of advisory committee assignments with 510(k) recall rate, 2003–2009.

FIGURE C-15 Association of advisory committee assignments with 510(k) recall rate, 2003–2009.

  • Devices associated with certain advisory committee assignments were more likely to be affected by recall.

  • For example, Radiology (17.8% vs 10.1%), Cardiovascular (14.6% vs 11.2%), Anesthesiology (6.8% vs 5.4%), and Clinical Chemistry (8.6% vs 5.9%) represented a higher percentage of devices among recalled than unaffected 510(k)s.

  • In contrast, Dental (1.8% vs 8.9%), Ear, Nose & Throat (0.3% vs 1.1%), and General & Plastic Surgery (8.4% vs 12.3%) represented a lower percentage among recalled devices compared with devices unaffected by recall.

Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×

TABLE C-3 Medical Device Reporting and 510(k)s

Description

Number

510(k)s with MDR

7823

Total MDRs

182,394

MDRs with Death

2361

MDRs with Injury

53,879

MDRs with Malfunction

12,110

MDRs Other

5051

  • 182,394 MDRs associated with 7,823 510(k)s from 1996 to 2009 have been filed.

  • Nearly two-thirds of the reports (66.4%) are associated with a device malfunction.

  • 29.5% of the reports are associated with a patient injury, although <2% involve a patient death.

  • Data on MDRs are included in this report because they were specifically requested by the Institute of Medicine. These data are subject to incomplete reporting, insufficient information, and misclassifications and should be interpreted with caution.

Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
FIGURE C-16 Medical device reporting and recalls.

FIGURE C-16 Medical device reporting and recalls.

  • 41.8% of 510(k) MDRs were associated with 510(k)s that were subject to recall.

  • 30.0% of death adverse event reports, 34.0% of reported patient injuries, and 45.6% of device malfunction reports involving 510(k)s were associated with recalled 510(k)s.

  • Data on MDRs are included in this report because they were specifically requested by the Institute of Medicine. These data are subject to incomplete reporting, insufficient information, and misclassifications and should be interpreted with caution.

SUMMARY OF KEY FINDINGS

  • 48,402 510(k)s were cleared by FDA between January 1, 1996, and December 31, 2009, and were available for analysis.

  • From 2003 to 2009, 3,132 unique 510(k)s were subject to recall. Among 510(k)s affected by recall, 73.9% were recalled a single time and 26.1% were recalled more than once, including nearly 2% that were recalled more than 5 times.

  • Among 510(k)s cleared in 2003–2009, 98.4% remained free of recall 1 year following the decision. Longer term follow-up shows that 92.6% and 91.5% of 510(k)s remain free of recall 5 and 6 years, respectively, following regulatory clearance.

  • The annual 510(k) recall rate is highest in the first 3 years following clearance (1.6–1.9%/year). Lower recall rates are observed in years 5 and 6 post clearance (0.9–1.1%/year).

  • More than half the 510(k) recalls are due to manufacturing process errors (28.8%) or device design issues (28.4%). Materials and com-

Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×

ponent issues (16.3%) and change control processes (11.9%) account for the majority of the remaining 510(k) recalls.

  • Compared to 510(k)s unaffected by recall, recalled 510(k)s are more likely to have been reviewed by a third party or submitted as a Special application (rather than Traditional or Abbreviated). Recalls are also more likely to affect 510(k)s involving life sustaining devices and Class III devices.

Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×

This page intentionally left blank.

Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
Page 73
Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
Page 74
Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
Page 75
Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
Page 76
Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
Page 77
Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
Page 78
Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
Page 79
Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
Page 80
Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
Page 81
Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
Page 82
Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
Page 83
Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
Page 84
Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
Page 85
Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
Page 86
Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
Page 87
Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
Page 88
Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
Page 89
Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
Page 90
Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
Page 91
Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
Page 92
Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
Page 93
Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
Page 94
Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
Page 95
Suggested Citation:"Appendix C: 510(k) Premarket Notification Analysis of FDA Recall Data." Institute of Medicine. 2011. Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report. Washington, DC: The National Academies Press. doi: 10.17226/13020.
×
Page 96
Next: Appendix D: Trustworthy Medical Device Software »
Public Health Effectiveness of the FDA 510(k) Clearance Process: Measuring Postmarket Performance and Other Select Topics: Workshop Report Get This Book
×
Buy Paperback | $45.00 Buy Ebook | $35.99
MyNAP members save 10% online.
Login or Register to save!
Download Free PDF

The Food and Drug Administration (FDA) is responsible for ensuring that medical devices are safe and effective before they go on the market. Section 510(k) of the Federal Food, Drug, and Cosmetic Act requires a manufacturer of medical devices to notify FDA of its intent to market a medical device at least 90 days in advance. That window of time allows FDA to evaluate whether the device is substantially equivalent to a product already legally on the market (called a predicate), in which case the device does not need to go through the premarket approval (PMA) process.

As part of its assessment of the FDA's premarket clearance process for medical devices, the Institute of Medicine (IOM) held a workshop on July 28, 2010 to discuss how medical devices are monitored for safety after they are available to consumers. Its primary focus was on monitoring the safety of marketed medical devices, including FDA's postmarket surveillance activities, analysis of safety concerns that resulted in medical device recalls, and non-FDA sources of adverse-event information.

Public Health Effectiveness of the FDA 501(K) Clearance Process summarizes the views of the workshop participants.

  1. ×

    Welcome to OpenBook!

    You're looking at OpenBook, NAP.edu's online reading room since 1999. Based on feedback from you, our users, we've made some improvements that make it easier than ever to read thousands of publications on our website.

    Do you want to take a quick tour of the OpenBook's features?

    No Thanks Take a Tour »
  2. ×

    Show this book's table of contents, where you can jump to any chapter by name.

    « Back Next »
  3. ×

    ...or use these buttons to go back to the previous chapter or skip to the next one.

    « Back Next »
  4. ×

    Jump up to the previous page or down to the next one. Also, you can type in a page number and press Enter to go directly to that page in the book.

    « Back Next »
  5. ×

    Switch between the Original Pages, where you can read the report as it appeared in print, and Text Pages for the web version, where you can highlight and search the text.

    « Back Next »
  6. ×

    To search the entire text of this book, type in your search term here and press Enter.

    « Back Next »
  7. ×

    Share a link to this book page on your preferred social network or via email.

    « Back Next »
  8. ×

    View our suggested citation for this chapter.

    « Back Next »
  9. ×

    Ready to take your reading offline? Click here to buy this book in print or download it as a free PDF, if available.

    « Back Next »
Stay Connected!