... Emerging Technologies for Nutrition Research, 1997 Pp.
From page 298...
... heart-rate recording (Spurr et al., 1988) , pedometers and accelerometers (Kashiwazaki et al., 1986; Montoye, et al., 1983)
From page 299...
... and integration of ventilation and gas concentrations and, thus, the calculation of oxygen uptake. TEEM 100 Specifications The TEEM 100 is the newest of the portable systems and has been available commercially for approximately 4 years (TEEM 100 Operator's Manual, 1993)
From page 301...
... TABLE 13-2 COSMED K2 Specifications Weight: 0.85 kg (transmitter + battery pack) VE: Turbine flowmeter in facemask (range: 2 to > 300 liters· min-1)
From page 302...
... VO2 = VE × F1O2 -FeO2) , (Equation 13-1)
From page 303...
... TABLE 13-3 Average Values of Minute Ventilation (liters · min-1, BTPS) Intensity COSMED1 COSMED2 DB Stage 2 79 ± 19 79 ± 19 78 ± 19 Stage 4 99 ± 24 98 ± 26 97 ± 25 Stage 6 121 ± 33 119 ± 32 122 ± 33 Maximal 151 ± 35 151 ± 32 159 ± 40 NOTE: BTPS, body temperature and pressure, saturated gas; DB, Douglas bag.
From page 304...
... Mean values of VO2 among systems showed significant differences only at Stage 6 of the submaximal exercise test (Table 13-5)
From page 305...
... TABLE 13-6 Oxylog Specifications Weight: < 2 kg VI: Turbine flowmeter in facemask (range: 2–150 liters · min-1) FIO2 and FeO2: Figaro fuel cell (range: 0–25%)
From page 306...
... highest intensity (4.9% lower for Oxylog) , while Oxylog values for VO2 were 4.4 percent lower (p < 0.01)
From page 307...
... TABLE 13-8 Oxygen uptake, liters · min-1 (mean ± SE) Grade (%)
From page 308...
... carriage, lifting, lift and carry, and obstacle course navigation) in chemical protective clothing.
From page 309...
... References Ballal, M.A., and I.A. MacDonald 1982 An evaluation of the Oxylog as a portable device with which to measure oxygen consumption.
From page 310...
... Oxylog Operator's Manual 1994 Operator's Manual. Oxylog 2.
From page 311...
... JOHN PATTON: Advantage between them?
From page 312...
... JOHN PATTON: Yes, and barometric pressure. That is all you are going to need.
From page 313...
... DOUGLAS WILMORE: So thermal loads and things [affect the results]
From page 314...
... apply it to the equation for oxygen uptake, there is going to be little change, little effect on oxygen uptake per se. I mean with just the masked condition itself, you tend to hypoventilate a little bit, but you extract more oxygen, so when you make the calculation for oxygen uptake there is little change.