Skip to main content

Currently Skimming:


Pages 31-67

The Chapter Skim interface presents what we've algorithmically identified as the most significant single chunk of text within every page in the chapter.
Select key terms on the right to highlight them within pages of the chapter.


From page 31...
... NCHRP 07‐19(02) Final Report 31 Chapter 3: Combined Findings from Phases 1 and 2 A major component of the research involved field testing a variety of commercially available pedestrian and bicycle counting technologies by comparing the counts produced by the technologies with manual counts taken from video footage.
From page 32...
... Figure 3‐ The graph example, adjusted intended multiuse tubes ove Figure 3‐ 1.   Exam ical analysi the pneuma after the firs for a mixed path site. Pr rcounted, bu 2.   Befor Adjus ple of a Gra s also shows tic tubes on t two days o traffic situat ior to the adj t overall cou e and After  tment  phical Displ patterns in the Midtown f data collect ion, which re ustment, the nt accuracy Comparison 32 ay of Count  the data in te Greenway i ion during P quires diffe tubes unde improved.
From page 33...
... NCHRP 07‐19(02) Final Report 33 Accuracy Calculations  Four accuracy and precision measures were calculated for each technology.
From page 34...
... NCHRP 07‐19(02) Final Report 34 By taking the absolute values, over and undercounts of the same magnitude no longer balance each other out, but rather both count toward the total accuracy.
From page 35...
... NCHRP 07‐19(02) Final Report 35 ߤ௜ ൌ ܣ௜݁ఈ೔ାࢼࢄ࢏ ൌ ܣ௜݁ఈ೔Π௞݁ఉೖ௑ೖ೔ That is, the rate is assumed to vary linearly with the automated count.
From page 36...
... overcoun eliminate Based on motor ve understa character the detec passes th either mo where th In additio repeater the repea Accuracy In Figure count of b truth to b encroach apparent One outli automate would be removing (a)  Bicy Figure 3‐ t, when in fa these error this experie hicles are lik nding of the istics of the tion zone.
From page 37...
... NCHRP 07‐19(02) Final Report 37 Table 3‐1.  Accuracy and Consistency Values for Radar Sensor  Dataset  APD  AAPD  WAPD  r  N  Average  Hourly  Volume  Bicyclists  256%  261%  18%  0.89  32  70  Bicyclists  (no outlier)
From page 38...
... the direct the data c than the b bounding sources o differenti Accuracy The therm The obse r=0.912 w Figure 3‐ Regressi Based on with the c there app multiplic develope AIC, and h ion specified ollector to m icycle or pe boxes that w f error. Futu ate bicyclist and Consis al imaging rved accurac ith a sampl 4.  Accur on Correctio visual inspe aveat that th ears to be li ative factor.
From page 39...
... NCHRP 07‐19(02) Final Report 39 Table 3‐2.  Correction Functions for Active Infrared Sensor  Intercept  Night  Rain  AIC  ‐0.0263 (0.163)
From page 40...
... NCHRP 07‐19(02) Final Report 40 sinking a wooden post into the ground adjacent to the facility being counted, as is common when used as part of a combination sensor unit.
From page 41...
... Figure 3‐ The fact t technolog installatio However counts ov obtained correctio needed to correctio Device‐Sp Evaluatin these tech calculate 5.  Accur hat the prod y plays an im n are impor , as can be in er a variety from any of n factor need adjust the o n factors. ecific Accur g accuracy a nologies, bu separate fac (a)
From page 42...
... NCHRP 07‐19(02) Final Report 42 Table 3‐3.   Accuracy and Consistency Metrics on a Site‐ and Device‐Specific Basis for  Passive Infrared Sensors  Site  Product  APD  AAPD  WAPD  r  N  Average Hourly  Volume  Overall  Average  ‐  −10%  17% −11%  0.96  392  261  Overall  Product A  A −1%  12% −4%  0.98  238  284  L St  A −2%  10% −3%  0.94  39  537  Four Mile  Run  A  2%  8%  3%  0.90  42  171  Fell Street  A −5%  10% −3%  0.97  26  43  15th  Avenue  A −5%  5% −4%  0.99  16  367  Key Bridge  A −12%  13% −13%  0.99  48  393  Berkeley  A −14%  14%  14%  0.96  6  151  Sycamore  A  14%  15%  12%  0.99  24  160  Loyola  A −18%  19% −18%  0.86  12  64  Overall  Product B  B −26%  26% −27%  0.98  115  263  Midtown  Greenway  B −26%  26% −27%  0.97  24  436  Four Mile  Run  B −21%  22% −21%  0.91  42  171  Key Bridge  B −29%  29% −30%  0.98  31  351  Sycamore  B −33%  33% −31%  0.85  18  96  Overall  Product C  C −14%  14% −14%  0.99  39  113  Four Mile  Run  C −14%  14% −14%  0.99  39  113  Notes:  APD = average percentage deviation, AAPD = average of the absolute percent difference, WAPD = weighted average percentage  deviation, r = Pearson's Correlation Coefficient, N = number of detectors, Average volume = hourly average pedestrian and bicycle  counts based on video observation.  Correction Functions The various correction function forms that were tested for passive infrared counters are presented in Table 3‐4.
From page 43...
... NCHRP 07‐19(02) Final Report 43 greater degrees of undercounting, as the adjustment factor is positive.
From page 44...
... Worse pe distinguis There is m 3‐6, there than in th temperat problem most of th difficult t On the ot at very hi very cold attributed current r enough c recomme winter te Figure 3‐ Worse pe The data Figure 3‐ rformance at hing people f inimal evid are limited e overall da ures above 9 is that high t e data in th o conclusive her hand, th gh volumes. temperatur to heavily i esearch beca lothing to ha nded that pr mperatures 6.  Passiv rformance in collected by 7.
From page 45...
... Figure 3‐ However observed detection However undercou The testin whether operate b Active I Qualitati The testin device ap with prev The activ transmitt 7.  Accur Range , during peri . On the surf s on passive , in the regre nting, as inc g was not a snow affects y detecting b nfrared  ve Experien g only inclu peared to fu ious experie e infrared se er and recei acy Compar   ods of rain, n ace, this wou infrared sen ssion analys luding the ex ble to captur device perfo ody heat.
From page 46...
... Accuracy The activ 3‐8. In pa WAPD = ‐ hours of d Figure 3‐ Regressi Based on again wit finding, th simple m were test fit was w Table 3‐5 In 0.079 0.0 0.079 0.09 Notes:   AI a 9 Ni and Consis e infrared se rticular, volu 7.62%, r = 0 ata.
From page 47...
... NCHRP 07‐19(02) Final Report 47 Effects of Different Conditions Because only one active infrared sensor was tested in the research, there was not enough variation in the data to ascertain whether any site‐level factors have an impact on the accuracy.
From page 48...
... two sets o tubes we longer pr inaccurat occasiona However counting. overall da Figure 3‐ Accuracy Figure 3‐ Figure 3‐ C with bic and site, a f pneumatic re damaged operly funct e counts is t lly occluded , this does no The data fro ta plot in Fi 9.  Accur and Consis 10 plots accu 11 plots accu ycle‐specifi nd Table 3‐ tubes, one i during the p ional when t hat 15th Ave the camera t seem like t m these sen gure 3‐10)
From page 49...
... Figure 3‐ The pneu counter t work. It c accuracy Universit were test tubes we sometime tubes or n crossed c 10.  Accur matic tube d echnologies.
From page 50...
... Figure 3‐11.  Accuracy Plot for Pneumatic  50 Tubes by Pr N oduct  CHRP 07‐19(02) Final Report
From page 51...
... Figure 3‐12.  Accuracy Plot for Pneumatic  51 Tubes by Pr N oduct and S CHRP 07‐19 ite  (02) Final Report
From page 52...
... NCHRP 07‐19(02) Final Report 52 Table 3‐6.  Accuracy and Consistency Values for Pneumatic Tubes by Product and Site  Site  Product  APD  AAPD  WAPD  r  N  Average  Volume  Overall Average ‐  −14.0%  16.6%  −14.1%  0.986  259  166  Bicycle‐specific  tubes −14.0%  16.6%  −14.0%  0.986  242  173  Overall Product A  A  −9.5%  10.8%  −11.1%  0.993  172  203  University  A  0.5%  1.2%  0.6%  0.998  17  206  Key Bridge  A  −11.1%  11.3%  −12.4%  0.981  48  97  L Street  A  −6.7%  10.8%  −10.4%  0.991  45  62  Fell Street  A  −20.2%  20.2%  −23.1%  0.987  23  248  Clarendon  A  −21.1%  21.1%  −22.9%  0.999  3  23  Milton  A  −7.7%  7.7%  −9.2%  0.997  36  503  Overall Product B  B  −52.6%  52.6%  39.9%  0.970  28  99  University  B  −34.3%  34.3%  −34.8%  0.805  11  199  Eastbank  B  −64.4%  64.4%  59.4%  0.781  17   34  Overall Product C  C  −7.3%  16.6%  −9.4%  0.920  43  76  Four Mile  C  8.9%  9.2%  7.6%  0.996  22  75  15th Street NW  C  −24.2%  24.2%  −26.6%  0.976  20  81  Oakland  C  −28.6%  28.6%  −28.6% 1  14  Standard Tubes  C  −15.2%  17.1%  −17.9%  0.936  17  62  Oakland  C  −15.2%  17.1%  −17.9%  0.936  17  62  Notes:  APD = average percentage deviation, AAPD = average of the absolute percent difference, WAPD = weighted average percentage  deviation,  r = Pearson's Correlation Coefficient, N = number of detectors, Average volume = hourly average pedestrian and bicycle  counts based on video observation.  Effects of Environmental Conditions Frozen tubes The research team hypothesized that pneumatic tubes will become less sensitive in very cold temperatures, as the rubber in the tubes hardens.
From page 53...
... NCHRP 07‐19(02) Final Report 53 specific tubes was left installed on the Midtown Greenway for the duration of the study (~5 months)
From page 54...
... NCHRP 07‐19(02) Final Report 54 Radio Beam  Qualitative Experience The research team faced significant difficulties evaluating the accuracy of one of the radio beam sensors.
From page 55...
... Figure 3‐   Table 3‐8 Dataset  Overall  Average  Four Mil Midtown 5th Aven Notes:   AP de Effects of The prim as is the c lighting, a variation radio bea between 13.  Accur .  Accur Pro e Run    ue  D = average perc viation, r = Pearso Environme ary hypothe ase with all nd rain wer in the weath m sensors in modes. acy Plot for  acy and Con duct  AP ‐  −10 B  −6.
From page 56...
... Inductiv Qualitati Inductive Both perm When eva addresse always co which is l and insta eliminate First, con 14. The lo is located located in maneuve the loop s Fell Stree also be co Figure 3‐ Second, c Minneapo zones.
From page 57...
... frequentl estimatin difficult t Finally, co Arlington The path path. Hen Figure 3‐ Accuracy Figure 3‐ volume a plots rep Midtown of nearly lower vol volume a slightly a y at high vol g the accura o determine nsider the e , Virginia, sh has a constr ce, this site 15.  Induc and Consis 16 compares nd the overa resent two s Greenway d 800 bicyclis umes, the fa ppears to inc ffected.
From page 58...
... Figure 3‐ Table 3‐9 (i.e., cons table, ind substanti Table 3‐9 Notes:  AP de Dataset  All data  All surfa Loyola  Universit All embe Sycamor Clarendo Key  Fell  (a)  Detect 16.  Accur provides the idering only uctive loop s al difference .
From page 59...
... NCHRP 07‐19(02) Final Report 59 Table 3‐10 presents accuracy and consistency statistics for inductive loops when considering facility volumes (i.e., including bypasses)
From page 60...
... NCHRP 07‐19(02) Final Report 60 For these installations, site‐specific correction factors should be developed to account for bypass errors and to improve the estimation of facility volumes.
From page 61...
... Figure 3‐ Both of th consisten are some are prom Table 3‐1 Product  Overall  Product  Product  Notes:  AP de co Effects of While the precipita condition Regressi As noted are neede tested bu 17.  Accur e piezoelect cy over the r what limited ising for this 1. Accur AP −4.0 A  −3.4 B  −5.2 D = average perc viation, r = Pearso unts based on vid Environme re was limit tion on accur s.
From page 62...
... NCHRP 07‐19(02) Final Report 62 have an intercept value of 0.0415 (95% CI: 0.024, 0.059)
From page 63...
... Figure 3‐ Figure 3‐ The insta due to oc 18.   Accur 19.  Accur Count llation at the clusion effec acy Plot for  acy and Con ers Compar Key Bridge ts) with a hi Pedestrian  sistency Plo ing Two Pro site, on the o gh rate of co 63 Volumes Es ts for Pede ducts  ther hand, s nsistency, as N timated fro strian Volum hows net un indicated in CHRP 07‐19 m Combina es from Co dercounting Table 3‐12.
From page 64...
... NCHRP 07‐19(02) Final Report 64 Table 3‐12.  Accuracy and Consistency Metrics for Pedestrian Volumes from Combination  Counters  Regression Corrections Table 3‐13 shows the correction functions developed for pedestrian volume estimates using combination counters.
From page 65...
... NCHRP 07‐19(02) Final Report 65 SUMMARY  Table 3‐14 provides a combined comparison of accuracy and consistency values by site, product, and counting technology.
From page 66...
... NCHRP 07‐19(02) Final Report 66 Technology  Subset  APD  AAPD  WAPD  r  N  Average  Hourly  Volume  Inductive Loops  All data  0.6%  8.9%  7.6%  0.994  108  128  Surface loops  0.3%  7.6%  5.7%  0.997  29  145  Embedded loops  0.6%  9.4%  8.4%  0.993  79  122  Sycamore  7.9%  8.2%  7.7%  0.973  18  81  Loyola  7.9%  10.8%  9.8%  0.974  12  51  Clarendon −27.4%  27.4%  29.7%  1.000  3  25  Key  8.0%  8.1%  6.0%  0.996  31  92  University −5.0%  5.3%  5.0%  0.988  17  211  Fell −9.5%  9.5%  9.5%  0.998  27  194  All data*
From page 67...
... NCHRP 07‐19(02) Final Report 67 for the technologies tested.

Key Terms



This material may be derived from roughly machine-read images, and so is provided only to facilitate research.
More information on Chapter Skim is available.